Description of device functions Proline Prowirl 73
110 Endress+Hauser
11.6 OPERATION
Function description, OPERATION group
LANGUAGE Description
For selecting the language in which all messages are shown on the local display.
If you press the P keys simultaneously at startup, the language defaults to "ENGLISH".
Options with standard display:
ENGLISH
DEUTSCH
FRANCAIS
ESPANOL
ITALIANO
NEDERLANDS
NORSK
SVENSKA
SUOMI
PORTUGUES
POLSKI
CESKI
Available in addition with the graphical display option:
CHINESE
JAPANESE
RUSSIAN
Factory setting
Depends on country ä 180
ACCESS CODE Description
All data of the measuring system are protected against inadvertent change. Programming
is disabled and the settings cannot be changed until a code is entered in this function. If
the
P
keys are pressed in any function, the measuring system automatically goes to the
function. If programming is locked, a prompt to enter a code appears on the display.
You can enable programming by entering the private code (factory setting = 73,
ä 110, DEFINE PRIVATE CODE function).
• The programming levels are disabled if the operating elements are not pressed within
60 seconds following a return to the HOME position.
• Programming can also be disabled by entering any number (other than the private
code) in this function.
• Your Endress+Hauser representative can be of assistance if you mislay your private
code.
User input
Max. 4-digit number: 0 to 9999
DEFINE PRIVATE CODE Description
Use this function to specify the private code for enabling programming.
• Programming is always enabled if the code defined = 0.
• Programming has to be enabled before this code can be changed. When programming
is disabled this function cannot be edited, thus preventing others from accessing your
personal code.
User input
Max. 4-digit number: 0 to 9999
Factory setting
73
STATUS ACCESS Description
The access status for the function matrix appears on the display.
Display
ACCESS CUSTOMER (parameters can be modified)
LOCKED (parameters cannot be modified)