9 Trouble-shooting Promag 53 PROFIBUS-DP/-PA
116 Endress+Hauser
Amplifier EEPROM
data inconsistent
S MP SW-EEPROM
# 012
12 device failure BAD
0x0F
constant
Cause:
System error. Error when accessing
data of the measuring amplifier
EEPROM
Remedy:
Execute a warm start (= start up
the measuring system without net-
work interruption).
• PROFIBUS (Commuwin II):
Manufacturer-specific trans-
ducer block
→ Device matrix
(V8H5)
• Local display:
SUPERVISION
→ SYSTEM →
OPERATION
→ SYSTEM RESET
(
→ RESTART SYSTEM)
S-DAT failure /
S-DAT not inserted
S SENSOR HW DAT
# 031
31 sensor failure BAD
0x10
no limits
Cause:
System error
1. S-DAT™ is defective.
2. S-DAT™ is not plugged into I/O
board (missing)
Remedy:
1. Replace the S-DAT™.
Spare parts
→ Page 122
Check the spare part set num-
ber to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
2. Plug S-DAT™ into I/O-board
→
Page 124, 126
S-DAT data
inconsistent
S SENSOR SW DAT
# 032
32 sensor failure BAD
0x10
no limits
Cause:
System error. Error accessing the
calibration values stored in the S-
DAT™.
Remedy:
1. Check whether the
S-DAT™ is correctly plugged
into the amplifier board
→
Page 124, 126
2. Replace the S-DAT™ if it is
defective.
Spare parts
→ Page 122
Check the spare part set num-
ber to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
3. Replace measuring electronics
boards if necessary.
Spare parts
→ Page 122