Promag 53 PROFIBUS-DP/-PA 9 Trouble-shooting
Endress+Hauser 117
T-DAT failure S TRANSM. HW DAT
# 041
41 device failure BAD
0x0F
constant
Cause:
System error
1. T-DAT™ is faulty
2. T-DAT™ is not plugged into I/O
board (missing).
Remedy:
1. Replace the T-DAT™.
Spare parts
→ Page 122
Check the spare part set num-
ber to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
2. Plug T-DAT™ into I/O-board
→
Page 124, 126
T-DAT d a ta
inconsistent
S TRANSM. SW DAT
# 042
42 device failure BAD
0x0F
constant
Cause:
System error. Error accessing the
calibration values stored in the T-
DAT™.
Remedy:
1. Check whether the T-DAT™ is
correctly plugged into the I/O
board
→ Page 124, 126
2. Replace T-DAT™ if faulty.
Spare parts
→ Page 122
Check the spare part set num-
ber to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
3. Replace measuring electronics
boards if necessary.
Spare parts
→ Page 122
TOT could not be
restarted
S CHEKSUM. TOTAL.
# 111
111 device failure BAD
0x0F
constant
Cause:
System error. Totalizer checksum
error
Remedy:
1. Restart the device
2. Replace the amplifier board if
necessary.
Spare parts
→ Page 122
Save to T-DAT failed S LOAD T-DAT
# 205
205 device failure BAD
0x0F
constant
Cause:
System error. Data backup to
T-DAT failed, or error when acces-
sing the calibration values stored in
the T-DAT™.
Remedy:
1. Check whether the T-DAT™ is
correctly plugged into the I/O
board
→ Page 124, 126
2. Replace T-DAT™ if faulty.
Spare parts
→ Page 122
Check the spare part set num-
ber to ensure that the new,
replacement DAT is compatible
with the measuring electronics.
3. Replace measuring electronics
boards if necessary.
Spare parts
→ Page 122
Device status
message
Diagnosis message
(Control system)
Device status
message
(Display)
No. Initial status
Analog Input Block/
Totalizer Block
Measured
value Q /
Substatus/
alarm limit
Cause/remedy