Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 133
Operating conditions
Installation
Installation instructions Any orientation (vertical, horizontal)
Restrictions and additional installation instructions
→ Page 13 ff.
Inlet and outlet runs Inlet run: typically
≥ 5 x DN
Outlet run: typically
≥ 2 x DN
Length of connecting
cable
Permissible cable length L
max
depends on the fluid conductivity. A minimum
conductivity of 20
µS/cm is required for measuring demineralised water.
Permissible range = grey shaded area
Environment
Ambient temperature
−20...+60 °C (sensor, transmitter)
Note the following points:
• Install the device at a shady location. Avoid direct sunlight, particularly in
warm climatic regions.
• If both fluid and ambient temperatures are high, install the transmitter at a
remote location from the sensor (
→ “Fluid temperature range”).
Storage temperature –10...+50 °C (preferably +20 °C)
Degree of protection • Standard: IP 67 (NEMA 4X) for transmitter and sensor
• Optional: IP 68 (NEMA 6P) for remote version of Promag W and P sensor
Resistance to
vibrations and shock
Acceleration up to 2 g in accordance with IEC 68-2-6
(High-temperature version: no corresponding information available)
CIP cleaning Promag W: not possible
Promag P: possible (note max. temperature)
Promag H: possible (note max. temperature)
SIP cleaning Promag W: not possible
Promag P: possible with PFA (note max. temperature)
Promag H: possible (note max. temperature)
Electromagnetic
compatibility (EMC)
To EN 61326 and NAMUR recommendations NE 21
200
100
5
10 100
L max
200
[m]
[ S/cm]µ
L max
F06-xxxxxxxx-05-xx-xx-xx-006