10 Technical data Promag 53 PROFIBUS-DP/-PA
164 Endress+Hauser
Process connections with aseptic gasket seal (DN 2...25 )
Weld nipple DIN 11850
1.4404 / 316L
5*H**-U***********
Sensor Pipe di G L H x B
DN [mm] Tube [mm] [mm] [mm] [mm]
2...8 14 x 2 10 14 23.3 60 x 42
15 20 x 2 16 20 23.3 60 x 42
25 (DIN) 30 x 2 26 30 23.3 70 x 52
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-011
Weld nipples ODT/SMS
1.4404 / 316L
5*H**-V***********
Sensor Pipe di G L H x B
DN [mm] Tube [mm] [mm] [mm] [mm]
2...8 12.7 x 1.65 9.4 12.7 16.1 60 x 42
15 19.1 x 1.65 15.8 19.1 16.1 60 x 42
1" (25 ANSI) 24.5 x 1.65 22.1 25.4 16.1 70 x 52
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-013
Clamp ISO 2852
1.4404 / 316L
5*H**-W***********
Sensor Pipe di G L H x B
DN [inch] Clamp [mm] [mm] [mm] [mm]
1" (25 ANSI) Tube 25.4 x 1.65
(ISO; 1")
22.6 50.5 44.5 70 x 52
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-023
Clamp DIN 32676
1.4404 / 316L
5*H**-0***********
Sensor Pipe di G L H x B
DN [mm] Clamp [mm] [mm] [mm] [mm]
2...8 Tube 14 x 2
(DIN 11850; DN 10)
10 34.0 41.0 60 x 42
15 Tube 20 x 2
(DIN 11850; DN 15)
16 34.0 41.0 60 x 42
25 (DIN) Tube 30 x 2
(DIN 11850; DN 25)
26 50.5 44.5 70 x 52
– Fitting length = (2 x L) + 86 mm
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 156)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-019