Promag 53 PROFIBUS-DP/-PA 10 Technical data
Endress+Hauser 169
Process connections with gasket seal (DN 40...100)
Weld nipples DIN 11850
1.4404 / 316L
5*H**-U***********
DN
[mm]
di
[mm]
G
[mm]
D
[mm]
L
[mm]
L1
[mm]
LK
[mm]
4038.04392421971.0
50 50.0 55 105 42 19 83.5
65 66.0 72 121 42 21 100.0
80 81.0 87 147 42 24 121.0
100 100.0 106 168 42 24 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-002
Weld nipples ODT
1.4404 / 316L
5*H**-V***********
DN
[mm]
di
[mm]
G
[mm]
D
[mm]
L
[mm]
L1
[mm]
LK
[mm]
4035.34092421971.0
50 48.1 55 105 42 19 83.5
65 59.9 66 121 42 21 100.0
80 72.6 79 147 42 24 121.0
100 97.5 104 168 42 24 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-002
Clamp ISO 2852
1.4404 / 316L
5*H**-W***********
DN
[mm]
di
[mm]
G
[mm]
D
[mm]
L
[mm]
LK
[mm]
40 35.6 50.5 92 68.5 71.0
50 48.6 64.0 105 68.5 83.5
65 60.3 77.5 121 68.5 100.0
80 72.9 91.0 147 68.5 121.0
100 97.6 119.0 168 68.5 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-005
Clamp DIN 32676
1.4404 / 316L
5*H**-0***********
DN
[mm]
di
[mm]
G
[mm]
D
[mm]
L
[mm]
LK
[mm]
40 38 50.5 92 61.5 71.0
50 50 64.0 105 61.5 83.5
65 66 91.0 121 68.0 100.0
80 81 106.0 147 68.0 121.0
100 100 119.0 168 68.0 141.5
– Fitting length = (2 x L) + 140 mm (DN 40...65) / + 200 mm (DN 80...100)
– If pigs are used for cleaning, it is essential to take the inside diameters of measuring tube (Page 158)
and process connection (di) into account.
F06-xxHxxxxx-06-09-07-xx-008