TOCII CA72TOC Troubleshooting
Endress+Hauser 127
Faulty
adjustment
(adjustment
fault 3)
The calibration line
slope is negative or
zero.
CO
2
concentration
measured for
standard 1 is higher
than that for
standard 2
Standard solutions,
container empty
Have the standard solutions been prepared properly?
Do the two standard solutions have the same concentration?
Have the standard solutions been swapped?
Are the standard canisters empty?
Faulty
adjustment
(adjustment
fault 4)
KP value is smaller
than 30 or greater
than 150
Standard solutions
Have the standard solutions been prepared properly?
Replace the standard solutions.
Biological growth in the standard canisters.
Replace the canisters.
Dilution option - feed rate of pump P4 deviates from the values
determined. Use the program module SERVICE/PUMPS/P1/P4
Replace pump hose to determine the feed rate for pump P4.
Faulty
adjustment
(adjustment
fault 5)
CO
2
concentration <
min. permitted CO
2
value.
(~ -9.4 % gas card
measuring range)
Is the IR detector OK?
Convey pure supply gas through the IR detector.
Go to the programming module PRG/TEST/INPUTS/ANALOG
INPUTS and check whether the IR detector displays a negative
offset.
Replace the IR detector if the frequency displayed has dropped
below 10000 Hz.
Message Cause Tests or remedial action