TOCII CA72TOC Commissioning
Endress+Hauser 53
18 Gas circuit pressure monitoring
0 = OK
1 = 70 % of max. permitted pressure
2 = > max. permitted pressure
19 Check CO
2
basic line (0 = off, 1 = on)
20 Adjustment error (0 = off, 1 = on)
21 Empty (unused byte)
22 Empty (unused byte)
23 0 = No valid measured value available
1 = Valid measured value available
2 = New measured value determined (displayed for approx. 4 seconds)
24 Separator
25 0 = Sample
1 = Standard is dosed
26 Flushing strip and separation chamber with supply water
27 0 = Error shutdown active, no power is supplied to any of the units powered
by the power relay
1 = Power supply active
28 0 = Standard C1 is dosed
1 = Standard C2 is dosed
If relay 1 (byte 25) is set to 1
29 Flushing sample conditioning
30 Only important for two-channel operation
0 = Sample is taken from sample channel 1
1 = Sample is taken from sample channel 2
31 Flushing with carrier gas
32 A 0-1-0 change indicates that the process of dosing the sample into the furnace
has been completed.
33 Separator
34 to 39 TOC measured value (mg/l)
1 decimal place for measuring range A and B
0 decimal places for measuring range C and D
40 Separator
41 to 46 Only for channel 2
TOC measured value (mg/l)
1 decimal place for measuring range A and B
0 decimal places for measuring range C and D