TOCII CA72TOC Maintenance
Endress+Hauser 73
a0014230
Fig. 36: Pump P2
Pump P2: Pumps sample from the separation chamber into the furnace.
Hose: (BK-BK) 0.76 mm (0.03")
1 Pump retainer 4 Adjusting screw
2 Pump head 5 Hose cassette
3 Sample hose (BK-BK) 0.76 mm (0.03") ID 6 Motor
Step Display Description
1
2
#
Warning!
There is a risk of infection when working
with wastewater! For this reason, wear
protective goggles, protective gloves and
protective clothing.
Press the [E] key.
PUMPS
CALIBRATION
SERVICE
CLEANING
FILTERS
P 1/4 REPLACE HOSE
PUMP 2 REPLACE HOSE
PUMP 3 REPLACE HOSE
CALIBRATE PUMP 2
REPLACE HOSE P2
CAUTION! LIQUID CAN LEAK OUT
WEAR PROTECTIVE CLOTHES
[E] TO CONFIRM