Maintenance TOCII CA72TOC
74 Endress+Hauser
3 The hose is drained automatically.
4 1. Open the cover of the separation
chamber.
2. Empty the separation chamber with
the syringe (part of the scope of
supply).
3. Press the [E] key.
5 In areas with a low amount of dust:
Grease the new hoses with a light coating
of silicone grease.
1. Release the hose at the injection unit
and at the separation chamber.
2. Disengage the hose cassette.
3. Change the hose.
4. Use the dosing nozzle of the injector
to expand the hose end.
5. Engage the hose cassette.
6. Press the [E] key.
6 1. Seal the separation chamber.
2. Connect the hose at the suction side.
3. Press the [E] key.
The "Adjust pump P2" service menu is
started.
Step Display Description
XX
REPLACE HOSE P2
[E] TO CONFIRM
PLEASE WAIT
PUMP CONVEYS BACKWARDS
REPLACE HOSE P2
[E] TO CONFIRM
EMPTY SEPARATION CHAMBER USING NOZZLE!
REPLACE HOSE P2
[E] TO CONFIRM
REPLACE PUMP HOSE
REPLACE HOSE P2
[E] TO CONFIRM
SHUT SEPARATION CHAMBERAND
ATTACH HOSE P2 ON SUCTION SIDE