EasyManua.ls Logo

ero electronic DPS - SECTION 5 INSTRUMENT CALIBRATION PROCEDURES; 5.1 Dip Switches Location

ero electronic DPS
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
43
SECTION 5 INSTRUMENT CALIBRATION PROCEDURES
5.1 DIP SWITCHES LOCATION
To start with the calibration procedure, the internal
DIP SWITCHES , mounted on CPU card, must be
positioned as shown at left:
NOTE: during calibration procedure the serial
communication interface will be disabled.
Fig. 5.1
SW1 = ON
SW2 = OFF
ON
2
1

Table of Contents