60
E401E401
E401E401
E401
The instrument detects the incorrect access to a protected memory
area (this error may be generated by a big noise only).
After 2 seconds the instrument restarts automatically.
Verify configuration and calibration parameters.
E402E402
E402E402
E402
This error may appear switch-on the instrument only.
The calibration and configuration parameters are not protected.
Go to configuration mode, select a parameter and push F pushbutton.
Return in operative mode.
If this error persists, send back the instrument to your supplier.
E403E403
E403E403
E403
The instrument detects a dip switch position changement or an
incorrect access to the program memory area (this error may be
generated by a big noise only).
After 2 seconds the instrument restarts automatically.
Verify configuration and calibration parameters.