58
E201E201
E201E201
E201
Incorrect calibration data in EAROM memory.
It may appear at instrument switching on in operative mode.
The instrument does not start to operate.
It remakes this check every 2 seconds.
If this error persists, remake the calibration procedure.
E211/E212E211/E212
E211/E212E211/E212
E211/E212
Autozero errors during calibration procedure.
The instrument measure an internal autozero value too negative
(E211) or too positive (E212).
It may appear during calibration procedure.
The instrument will not operate and after 6 seconds it will return to the
begining of the calibration procedure.
If this error persists, send back the instrument to your supplier.
E213E213
E213E213
E213
Too small input span.
It may appear during calibration procedure when the instrument
checks the input calibration and detects an input span less than 450
µA for current input or 110 mV for 5 V or 220 mV for 10 V or 1300 µV
for TC and mV input or 13 Ω for RTD input
Push F pushbutton and remake the calibration procedure.
If this error persists, send back the instrument to your supplier.
E214/E215E214/E215
E214/E215E214/E215
E214/E215
Could junction calibration errors.
The instrument measures a compensation temperature less than
-20 °C (E214) or higher than +70 °C (E215).
It may appear during calibration procedure.
Push F pushbutton and remake the calibration procedure.
If this error persists, send back the instrument to your supplier.
E216E216
E216E216
E216
Out of range during calibration.
The instrument detects an out of range signal or an open circuit
condition during input calibration.
The instrument shows the error message for 6 seconds then it will
return to the calibration point where the error was found.
The new calibration of this parameter will be lost, verify the input
signal and redo the calibration procedure correctly.
E217E217
E217E217
E217
The instrument detects a too big difference between could junction
measurement and calibration set during could junction calibration.
The instrument shows the error message for 6 seconds then it will
return to the calibration point where the error was found.
The new calibration of this parameter will be lost.
Verify the input signal and redo the calibration procedure correctly.