57
7.4 ERROR DESCRIPTIONS7.4 ERROR DESCRIPTIONS
7.4 ERROR DESCRIPTIONS7.4 ERROR DESCRIPTIONS
7.4 ERROR DESCRIPTIONS
E001E001
E001E001
E001
The alarm threshold value are incompatible with the actual readout
range or their values in memory are incorrect.
It may appare at instrument switching on in operative mode.
The instrument does not start to operate.
Push contemporarily and pushbutton and force the threshold
values at the initial scale value.
Set the desired threshold values.
E038E038
E038E038
E038
EAROM memory reading error.
It may appear at instrument switching on in operative mode.
The instrument does not start to operate.
The instrument remakes this check every 2 seconds.
If this error persists, send back the instrument to your supplier.
E039E039
E039E039
E039
EAROM memory writing error during operative mode.
It may appear during operative mode.
This error message will be deleted automatically after 13 seconds.
The new values will be enabled but they will be lost at instrument
switch-off.
Send back the instrument to your supplier.
E101E101
E101E101
E101
Incorrect configuration data in EAROM memory.
It may appear at instrument switching on in operative mode.
The instrument does not start to operate.
It remakes this check every 2 seconds.
If this error persists, remake the configuration procedure.
E102E102
E102E102
E102
Inconsistent configuration parameters.
It may appear at instrument switching on in operative mode.
The instrument does not start to operate.
It remakes this check every 2 seconds.
If this error persists, remake the configuration procedure.
E138E138
E138E138
E138
EAROM reading error during configuration procedure.
It may appear during configuration procedure.
Push F pushbutton and restart the configuration procedure.
If this error persists, send back the instrument to your supplier.
E139E139
E139E139
E139
EAROM writing error during configuration procedure.
It may appear during configuration procedure.
Push F pushbutton and restart the configuration procedure.
If this error persists, send back the instrument to your supplier.