METALJET D2+ INSTALLATION
Revision 19, November 2016 Page 4-3
Figure 4-1. The x-ray head is semi-rigidly connected to the pump box by the supply high-pressure tubing and
the return bellow for the metal-jet loop. In addition, 6 mm outer diameter flexible hoses for cooling water
supply and return run from the opening in the pump box towards the x-ray head.
In order not to over-stretch or over-compress the metal-jet loop components during these adjustments, there
are limitations to their vertical positions in relation to each other, see Figure 4-2.
Minimum distance (pump-box lid to lower end of front cover): 45 cm
Maximum distance (pump-box lid to lower end of front cover): 55 cm
In addition, it is also important that the jet-exit port (see Section 4.3.2.2) is placed straight above the hole in
the pump-box lid, see Figure 4-2 and Figure 4-3.
High-pressure tubing
Flex-hose assembly