FM5000C - FM5000PLUS
Maternal/Fetal Monitor
Mean arterial pressure:
20mmHg~165mmHg(2.7kPa~22.0kPa)
Diastolic pressure
10mmHg~150mmHg(1.3kPa~20.0kPa)
Systolic pressure:
40mmHg~135mmHg(5.3kPa~18.0kPa)
Mean arterial pressure:
20mmHg~110mmHg(2.7kPa~14.7kPa)
Diastolic pressure
10mmHg~100mmHg(1.3kPa~13.3kPa)
±0.1kPa or ±1mmHg , whichever is the larger.
Time interval available for
auto measure mode
1min, 2min, 2.5min, 3min, 4min, 5min, 10min, 15min,
30min, 60min, 90min, 120min, 180min, 240min,
480min
(10) SpO
2
Measurement
range and accuracy
a) Analog SpO
2
module:
Measurement range: 0%~100%;
Accuracy: ±2% (measured without motion in adult/pediatric mode) or ±3%
(measured without motion in neonate mode) within the measurement range of
70%~100%.
b) MasimoSpO
2
module:
Measurement range: 1%~100%;
Accuracy: ±2% (measured without motion in adult/pediatric mode), ±3%
(measured with motion in adult/pediatric mode) or ±3% (measured without
motion in neonate mode) within the measurement range of 70%~100%.
c) NellcorSpO
2
module:
Measurement range: 0%~100%;
Accuracy: ±2% (measured without motion in adult/pediatric mode) or ±3%
(measured without motion in neonate mode) within the measurement range of
70%~100%.
d) The accuracy is not specified within0%~69%.