II-8
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
User Manual
Mean arterial pressure:
20mmHg165mmHg(2.7kPa22.0kPa)
Diastolic pressure
10mmHg150mmHg(1.3kPa20.0kPa)
Neonate
Systolic pressure:
40mmHg135mmHg(5.3kPa18.0kPa)
Mean arterial pressure:
20mmHg110mmHg(2.7kPa14.7kPa)
Diastolic pressure
10mmHg100mmHg(1.3kPa13.3kPa)
Accuracy
±0.1kPa or ±1mmHg , whichever is the larger.
Measuremode
Manual, auto, continuous
Time interval available for
auto measure mode
1min, 2min, 2.5min, 3min, 4min, 5min, 10min, 15min,
30min, 60min, 90min, 120min, 180min, 240min,
480min
Continuous
5min
(10) SpO
2
Item
Specification
Display resolution
1%
Measurement
range and accuracy
a) Analog SpO
2
module:
Measurement range: 0%~100%;
Accuracy: ±2% (measured without motion in adult/pediatric mode) or ±3%
(measured without motion in neonate mode) within the measurement range of
70%~100%.
b) MasimoSpO
2
module:
Measurement range: 1%~100%;
Accuracy: ±2% (measured without motion in adult/pediatric mode), ±3%
(measured with motion in adult/pediatric mode) or ±3% (measured without
motion in neonate mode) within the measurement range of 70%~100%.
c) NellcorSpO
2
module:
Measurement range: 0%~100%;
Accuracy: ±2% (measured without motion in adult/pediatric mode) or ±3%
(measured without motion in neonate mode) within the measurement range of
70%~100%.
d) The accuracy is not specified within0%~69%.