6 Mounting
FLUXUS F502TE 6.2 Transducers
31
UMFLUXUS_F502TEV1-5EN, 2020-05-29
Mounting
When measuring in diagonal arrangement, the transducer mounting fixtures are mounted on opposite sides of the pipe
(see Fig. 6.7). When measuring in reflection arrangement, the transducer mounting fixtures are mounted on the same side
of the pipe (see Fig. 6.8).
When measuring in diagonal arrangement with 2 beams in displaced X arrangement, 4 transducer mounting fixtures have
to be mounted. When measuring in reflection arrangement with a small transducer distance, only 1 transducer mounting
fixture has to be mounted (see Tab. 6.1).
In the following, the mounting of 2 transducer mounting fixtures in reflection arrangement is described (1 transducer
mounting fixture for each transducer).
Overview of the mounting steps
step 1
disassembly of the transducer mounting fixture Variofix L
step 2
fixation of the clasps to the tension straps
step 3
fixation of the tension strap to the pipe
step 4
fixation of the rail to the pipe
step 5
installation of the transducers in the mounting fixture Variofix L
Tab. 6.1: Approximate values for the mounting of both transducers in a Variofix L
transducer frequency (3rd character of the technical type) length of the rail [mm] transducer distance [mm]
F 368 < 94
G, H, K (****LI*) 368 < 94
G, H, K (except ****LI*) 348 < 89
M, P (Lamb wave transducers)
M, P (shear wave transducers)
234 < 84
< 100
Q 176 < 69
Fig. 6.7: Transducer mounting fixture Variofix L
(diagonal arrangement)
Fig. 6.8: Transducer mounting fixture Variofix L
(reflection arrangement)