6 Mounting
6.2 Transducers FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
34
Quick release clasp
• Insert the tension strap into the tension strap clamp and the metal spring (see Fig. 6.14 and Fig. 6.15).
• Position the clasp, the tension strap clamp and the metal spring on the pipe (see Fig. 6.13):
– on a horizontal pipe, mount the tension strap clamp on the side of the pipe, if possible
– mount the metal spring opposite to the tension strap clamp
• Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.15).
• Tighten the tension strap.
• Tighten the clasp screw.
Ratchet clasp
•
Insert the tension strap into the tension strap clamp and the metal spring (see Fig. 6.17).
The metal spring does not have to
be mounted on:
– steel pipes
– pipes with an outer pipe diameter < 80 mm
– pipes that are not subjected to significant temperature fluctuations
• Position the ratchet clasp, tension strap clamp and metal spring (if necessary) on the pipe (see Fig. 6.13):
– on a horizontal pipe, mount the tension strap clamp laterally to the pipe, if possible
– mount the metal spring (if necessary) opposite to the tension strap clamp
• Place the tension strap around the pipe and pass it through the slot of the clasp screw (see Fig. 6.18).
• Tighten the tension strap.
• Cut off the protruding tension strap (see Fig. 6.18).
• Tighten the clasp screw.
Fig. 6.14: Tension strap with clamp Fig. 6.15: Tension strap with quick release clasp
and metal spring
1 – tension strap clamp 1 – clasp screw
2 – metal spring
Fig. 6.16: Tension strap with band clamp clasp
1 – clasp screw
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
1
2
1
1