6 Mounting
FLUXUS F502TE 6.2 Transducers
33
UMFLUXUS_F502TEV1-5EN, 2020-05-29
Ratchet clasp
Cut the tension strap to length (pipe circumference + at least 120 mm).
Insert approx. 100 mm of the tension strap into part (1) and (2) of the ratchet clasp (see Fig. 6.12 a).
Bend the tension strap.
Insert the tension strap into part (1) of the ratchet clasp (see Fig. 6.12 b).
Tighten the tension strap.
Repeat the steps for the second tension strap.
Step 3: Fixation of the tension strap to the pipe
One tension strap is fixed to the pipe (see Fig. 6.13). The second tension strap is mounted later.
Select the installation instructions according to the supplied clasp:
Band clamp clasp
Insert the tension strap into the tension strap clamp (see Fig. 6.14).
Position the clasp and the tension strap clamp on the pipe (see Fig. 6.13). On a horizontal pipe, mount the tension strap
clamp laterally to the pipe, if possible.
Place the tension strap around the pipe and insert it into the clasp (see Fig. 6.16).
Tighten the tension strap.
Tighten the clasp screw.
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
Debur sharp edges.
Wear the required personal protective equipment.
Observe the applicable rules.
Fig. 6.12: Ratchet clasp with tension strap
Fig. 6.13: Tension strap with tension strap clamp and metal spring on the pipe
1 tension strap clamp
2 clasp
3 metal spring
a b
1 2 1
3
1
2