6 Mounting
6.2 Transducers FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
38
Repeat the steps to fix the second rail (see Fig. 6.26).
Step 5: Installation of the transducers in the mounting fixture Variofix L
Press the transducers firmly into their clamping fixtures in the covers until they are tightly fixed. The transducer cables
show in opposite directions (see Fig. 6.27).
Adjust the transducer distance displayed by the transmitter (see Fig. 6.28).
Fix the transducer cables with the strain relief clamp to protect them from mechanical strain (see Fig. 6.28).
Stick coupling foil (or apply some coupling compound for a short-term installation) on the contact surface of the transducers.
The coupling foil can be fixed to the contact surface with some coupling compound.
Put the covers with the transducers on the rails.
Correct the transducer distance, if necessary.
Fig. 6.25: Ratchet clasp with tension strap
1 sense of rotation
2 edge
3 lever
4 clasp screw with slot
Fig. 6.26: Pipe with 2 rails
Notice!
The arrows on the transducers and the covers have to point in the same direction.
Fig. 6.27: Installation of the transducers in the covers
1 cover
2 transducer clamping fixture
1
2
4
3
4
1
2