6 Mounting
FLUXUS F502TE 6.3 Temperature probe
43
UMFLUXUS_F502TEV1-5EN, 2020-05-29
6.3.2.2 Installation with FLEXIM clasp
Cut the tension strap to length (pipe circumference + at least 120 mm).
Insert approx. 20 mm of the tension strap into the slot of the clasp (see Fig. 6.37).
Bend the end of the tension strap.
Position the temperature probe on the pipe (see Fig. 6.36).
Place the tension strap around the temperature probe and the pipe.
Insert the tension strap through the parts (2) and (1) of the clasp.
Pull the tension strap firmly and engage it in the inner hooks of the clasp.
Tighten the screw of the clasp.
Fig. 6.35: Clasp
Fig. 6.36: Temperature probe on the pipe
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
Debur sharp edges.
Wear the required personal protective equipment.
Observe the applicable rules.
Fig. 6.37: FLEXIM clasp