6 Mounting
6.3 Temperature probe FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
44
6.3.2.3 Installation with quick release clasp
Cut the tension strap to length (pipe circumference + at least 120 mm).
Position the temperature probe on the pipe (see Fig. 6.36).
Place the tension strap around the temperature probe and the pipe.
Insert the tension strap into the clasp (see Fig. 6.38).
Tighten the tension strap.
Tighten the screw of the clasp.
6.3.3 Installation of the inline temperature probe
Screw the thermowell into the thread at the measuring point of the pipe (see Fig. 6.40). Seal the screw connection.
Insert the inline temperature probe as far as possible into the thermowell.
Fix the inline temperature probe using the clamping nut (see Fig. 6.39).
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
Debur sharp edges.
Wear the required personal protective equipment.
Observe the applicable rules.
Fig. 6.38: Quick release clasp
Fig. 6.39: Threaded thermowell and inline temperature probe
1 clamping nut
1
threaded thermowell inline temperature probe