6 Mounting
6.3 Temperature probe FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
44
6.3.2.3 Installation with quick release clasp
• Cut the tension strap to length (pipe circumference + at least 120 mm).
• Position the temperature probe on the pipe (see Fig. 6.36).
• Place the tension strap around the temperature probe and the pipe.
• Insert the tension strap into the clasp (see Fig. 6.38).
• Tighten the tension strap.
• Tighten the screw of the clasp.
6.3.3 Installation of the inline temperature probe
• Screw the thermowell into the thread at the measuring point of the pipe (see Fig. 6.40). Seal the screw connection.
• Insert the inline temperature probe as far as possible into the thermowell.
• Fix the inline temperature probe using the clamping nut (see Fig. 6.39).
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
Fig. 6.38: Quick release clasp
Fig. 6.39: Threaded thermowell and inline temperature probe
1 – clamping nut
1
threaded thermowell inline temperature probe