7 Connection
FLUXUS F502TE 7.4 Temperature probe
61
UMFLUXUS_F502TEV1-5EN, 2020-05-29
7.4.3 Connection with junction box
Connection of the extension cable to the transmitter
Remove the blind for the connection of the temperature probe (see Fig. 7.13).
Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.14).
Push the extension cable through the cap nut, the compression part, the basic part and the reducer.
Prepare the extension cable.
Shorten the external shield and brush it back over the compression part.
Insert the extension cable into the housing.
Screw the sealing ring side of the reducer into the transmitter housing.
Screw the basic part into the reducer.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the extension cable to the terminals of the transmitter (see Fig. 7.13 and Tab. 7.12).
Connection of the extension cable to the junction box
Remove the blind plug for the connection of the extension cable (see Fig. 7.17).
Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.16).
Push the extension cable through the cap nut and the compression part.
Prepare the extension cable.
Shorten the external shield and brush it back over the compression part.
Screw the sealing ring side of the reducer into the junction box.
Screw the basic part into the reducer.
Insert the extension cable into the junction box.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the extension cable to the terminals of the junction box (see Fig. 7.17 and Tab. 7.12).
Tab. 7.11: Terminal assignment (transmitter)
terminal extension cable
TF1, TX1 red
TF2, TX2 gray
TF3, TX3 blue
TF4, TX4 white
Fig. 7.16: Preparation of the extension cable
1 cap nut
2 compression part
3 basic part
4 reducer
5 external shield, brushed back
6 sealing ring side
110 mm
7 mm
1
2
3 4
6
5