7  Connection
FLUXUS F502TE  7.4  Temperature probe
61
UMFLUXUS_F502TEV1-5EN, 2020-05-29
7.4.3 Connection with junction box
Connection of the extension cable to the transmitter
• Remove the blind for the connection of the temperature probe (see Fig. 7.13).
• Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.14).
• Push the extension cable through the cap nut, the compression part, the basic part and the reducer. 
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Insert the extension cable into the housing.
• Screw the sealing ring side of the reducer into the transmitter housing.
• Screw the basic part into the reducer. 
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the transmitter (see Fig. 7.13 and Tab. 7.12).
Connection of the extension cable to the junction box
• Remove the blind plug for the connection of the extension cable (see Fig. 7.17).
• Open the cable gland of the extension cable. The compression part remains in the cap nut (see Fig. 7.16).
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the extension cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the junction box (see Fig. 7.17 and Tab. 7.12).
Tab. 7.11: Terminal assignment (transmitter)
terminal extension cable
TF1, TX1 red
TF2, TX2 gray
TF3, TX3 blue
TF4, TX4 white
Fig. 7.16: Preparation of the extension cable
1 – cap nut
2 – compression part
3 – basic part
4 – reducer
5 – external shield, brushed back
6 – sealing ring side
110 mm
7 mm
1
2
3 4
6
5