7 Connection
7.4 Temperature probe FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
62
Connection of the temperature probe to the junction box
• Remove the blind plug for the connection of the temperature probe (see Fig. 7.17).
• Open the cable gland of the temperature probe (see Fig. 7.16). The compression part remains in the cap nut.
• Push the cable of the temperature probe through the cap nut and the compression part.
• Prepare the cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the temperature probe to the terminals of the junction box (see Fig. 7.17 and Tab. 7.12).
Fig. 7.17: Junction box
1 – connection of the extension cable
2 – connection of the temperature probe
Tab. 7.12: Terminal assignment (junction box)
terminal extension cable (KL2) temperature probe (KL1)
1 red red
2 gray red/blue
3 white white
4 blue white/blue
1 2
KL2 KL1