7 Connection
7.4 Temperature probe FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
62
Connection of the temperature probe to the junction box
Remove the blind plug for the connection of the temperature probe (see Fig. 7.17).
Open the cable gland of the temperature probe (see Fig. 7.16). The compression part remains in the cap nut.
Push the cable of the temperature probe through the cap nut and the compression part.
Prepare the cable.
Shorten the external shield and brush it back over the compression part.
Screw the sealing ring side of the reducer into the junction box.
Screw the basic part into the reducer.
Insert the cable into the junction box.
Fix the cable gland by screwing the cap nut onto the basic part.
Connect the temperature probe to the terminals of the junction box (see Fig. 7.17 and Tab. 7.12).
Fig. 7.17: Junction box
1 connection of the extension cable
2 connection of the temperature probe
Tab. 7.12: Terminal assignment (junction box)
terminal extension cable (KL2) temperature probe (KL1)
1 red red
2 gray red/blue
3 white white
4 blue white/blue
1 2
KL2 KL1