7 Connection
FLUXUS F502TE 7.4 Temperature probe
63
UMFLUXUS_F502TEV1-5EN, 2020-05-29
7.4.4 Connection of the inline temperature probe
• Loosen the screws of the inline temperature probe cover and remove it (see Fig. 7.19).
• Prepare the cable (see Fig. 7.18).
• Cut off the shield and unused leads so that they are flush with the cable jacket. The shield and unused leads must not
have any contact with the cable gland.
• Loosen the cap nut of the cable gland and extract the compression part. The basic part remains in the housing of the
inline temperature probe.
• Insert the cable through the cap nut, the compression part and the basic part into the housing of the inline temperature
probe.
• Fix the cable gland by inserting the compression part into the basic part and by screwing the cap nut onto the basic part.
• Connect the cable to the terminals of the inline temperature probe (see Fig. 7.19 and Tab. 7.13 or the connection
scheme on the inside of the inline temperature probe cover).
Fig. 7.18: Preparation
1 – cap nut of cable gland
Fig. 7.19: Connecting head of the inline temperature probe
1 – cover
2 – screw
3 – basic part of cable gland
Tab. 7.13: Terminal assignment (inline temperature probe)
terminal extension cable
red red
red gray
white blue
white white
80 mm
7 mm
1
1
2
3
redred
whitewhite