7 Connection
7.4 Temperature probe FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
64
Connection to the transmitter
Remove the blind plug for the connection of the inline temperature probes (see Fig. 7.13).
Prepare the cables (see Fig. 7.20).
Open the cable gland.
Push the cables of the inline temperature probes through the cap nut, the compression part and the basic part.
Insert the cables into the housing.
Screw the sealing ring side of the basic part into the transmitter housing.
Brush the shields of the inline temperature probe cables back over the compression part. The shields and the basic part
of the cable gland must have direct contact.
Fix the cable gland by inserting the compression part into the basic part and by screwing the cap nut onto the basic part.
Connect the inline temperature probes to the terminals of the transmitter (see Fig. 7.13 and Tab. 7.14).
Fig. 7.20: Cable gland and preparation
1 compression part
2 shield
3 cap nut
4 basic part
Tab. 7.14: Terminal assignment (transmitter)
terminal extension cable
TF1, TX1 red
TF2, TX2 gray
TF3, TX3 blue
TF4, TX4 white
4
3
1
80 mm
7 mm
10 mm
2