9 Measurement
9.2 Measurement settings FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
80
9.2.3.1 Output of the measured value
Output range
When configuring an analog output, the output range has
to be defined.
Select the menu item Special Funct.\
SYSTEM settings\Proc. outputs\...\
I1 Output range.
Press ENTER.
Select a list item.
4/20 mA
other range
Press ENTER.
If other range is selected, enter the values Output MIN and Output MAX.
Press ENTER after each input.
Error output
In the following dialog, an error value can be defined which is to be transmitted if the source item cannot be measured,
e.g., if there are solids in the fluid.
Select a list item for the error output (see Tab. 9.2).
Press ENTER.
If Other value is selected, enter an error value. The value has to be within the output range.
Press ENTER.
Miscellaneous c-Medium sound speed of the fluid
SCNR ratio of useful signal and correlated noise signal
Signal signal amplitude of a measuring channel
VariAmp standard deviation of the signal amplitude
Density density of the fluid
Pressure pressure of the fluid
Tab. 9.1: Configuration of the outputs
source item list item output
Source item
Output range
Error-value
Active loop
Output Test
see annex A, p. 150
Special Funct.\SYSTEM settings\
Proc. outputs\...\I1 Output range
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Error-value
Notice!
The settings will be stored at the end of the dialog.