9 Measurement
FLUXUS F502TE 9.2 Measurement settings
81
UMFLUXUS_F502TEV1-5EN, 2020-05-29
Tab. 9.2: Error output
error value result
Minimum The lower limit of the output range is output.
Hold last value The last measured value is output.
Maximum The upper limit of the output range is output.
Other value The value has to be entered manually. It has to be within the limits of the output.
Example
source item:
output:
output range:
error delay:
volumetric flow rate
current output
4…20 mA
t
d
> 0
(see section 9.2.5 and Tab. 9.3)
The volumetric flow rate cannot be measured during the
time interval t
0
...t
1
. The error value will be output.
Tab. 9.3: Examples for the error output (output range: 4...20 mA)
list item output signal
Minimum (4.0 mA)
Hold last value
Maximum (20.0 mA)
v [m³/h]
t
0
t
1
t
I [mA]
20
4
t
d
t
0
t
1
t
I [mA]
20
4
t
0
t
1
t
I [mA]
20
4
t
d
t
0
t
1
t