9 Measurement
9.2 Measurement settings FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
82
Terminal assignment
The terminals for the connection of the output are displayed.
Press ENTER.
It will be displayed whether the current output is active or passive (here: active).
Output function test
The function of the output can now be tested.
Connect an external measuring instrument to the terminals of the installed output.
Enter a test value. It has to be within the output range.
Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
Select yes to repeat the test or no to return to the menu item SYSTEM settings.
Press ENTER.
9.2.4 Activation of an analog output
The measuring range of the source item has to be entered.
.
Other value
error value = 3.5 mA
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Active loop
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Output Test
Special Funct.\SYSTEM settings\Proc. outputs\...\I1= 10 mA\Again?
Tab. 9.3: Examples for the error output (output range: 4...20 mA)
list item output signal
I [mA]
20
4
t
d
t
0
t
1
t
Output Options
ENTER until
Current Loop
Meas.Values
Zero-Scale Val.
Full-Scale Val.
Test output
Error-val. delay
see annex A, p. 150
Notice!
An output can only be activated in the program branch
Output Options if it has previously been installed.