9 Measurement
9.2 Measurement settings FLUXUS F502TE
2020-05-29, UMFLUXUS_F502TEV1-5EN
82
Terminal assignment
The terminals for the connection of the output are displayed.
• Press ENTER.
It will be displayed whether the current output is active or passive (here: active).
Output function test
The function of the output can now be tested.
• Connect an external measuring instrument to the terminals of the installed output.
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select yes to repeat the test or no to return to the menu item SYSTEM settings.
• Press ENTER.
9.2.4 Activation of an analog output
The measuring range of the source item has to be entered.
.
Other value
error value = 3.5 mA
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Active loop
Special Funct.\SYSTEM settings\Proc. outputs\...\I1 Output Test
Special Funct.\SYSTEM settings\Proc. outputs\...\I1= 10 mA\Again?
Tab. 9.3: Examples for the error output (output range: 4...20 mA)
list item output signal
I [mA]
20
4
t
d
t
0
t
1
t
Output Options
ENTER until
Current Loop
Meas.Values
Zero-Scale Val.
Full-Scale Val.
Test output
Error-val. delay
see annex A, p. 150
Notice!
An output can only be activated in the program branch
Output Options if it has previously been installed.