9 Measurement
FLUXUS F502TE 9.2 Measurement settings
83
UMFLUXUS_F502TEV1-5EN, 2020-05-29
Press ENTER until Current Loop is displayed. Select yes to activate the output.
Press ENTER.
Measuring range
After an analog output has been activated in the program branch Output Options, the measuring range of the source
item has to be entered.
Select sign if the sign of the measured values is to be considered for the output.
Select absolut if the sign of the measured values is not to be considered for the output.
Press ENTER.
Enter the lowest expected measured value. The unit of measurement of the source item will be displayed.
Zero-Scale Val. is the value assigned to the value Output MIN of the output range.
Press ENTER.
Enter the highest expected measured value. The unit of measurement of the source item will be displayed.
Full-Scale Val. is the value assigned to the value Output MAX of the output range.
Press ENTER.
Function test
The function of the output can now be tested.
Connect an external measuring instrument to the terminals of the installed output.
Select yes to test the output.
Press ENTER.
Enter a test value for the selected physical quantity. If the external measuring instrument displays the entered value, the
output functions correctly.
Press ENTER.
Output Options\...\Current Loop
Output Options\...\Meas.Values
Output Options\...\Zero-Scale Val.
Output Options\...\Full-Scale Val.
Example
output: current output
output range: 4…20 mA
Zero-Scale Val.: 0 m³/h
Full-Scale Val.: 300 m³/h
volumetric flow rate = 0 m³/h, corresponds to 4 mA
volumetric flow rate = 300 m³/h, corresponds to 20 mA
Output Options\...\I1:Test output?
Output Options\...\I1:Test value