VisualSpreadsheet can determine the total volume it imaged with the following calculation:
Total Volume Imaged = Number of Images Taken during the Run × Volume per Picture (µm
3
)
VisualSpreadsheet can calculate PPM using either the Volume (ESD) or Volume (ABD) particle volume
measurements.
Total Volume of Particles using ESD or ABD (µm
3
)
PPM = ————————————————————— X 10
6
Total Volume Imaged (µm
3
)
Particle Count 1 x 10
12
(µm
3
)
P/mL = ———————————— X ———————
Total Volume Imaged (µm
3
) 1 (mL)
To locate the values used in the above calculations, from the Main window:
· For Height and Width, access Setup > Context > Capture (tab) > Acceptable Region (section) to
view the total number of pixels from left to right and from top to bottom.
· For Depth, access Setup > Context > Flow Cell (tab) > Flow cell (section) for the “Flow cell
depth” value (populated from the Flow Cell Database file).
· For total number of camera images, access Setup > Context > Run Summary (tab) > Images >
Total.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluid Imaging Technologies FlowCAM and is the answer not in the manual?

Fluid Imaging Technologies FlowCAM Specifications

General IconGeneral
BrandFluid Imaging Technologies
ModelFlowCAM
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals