- 67 -
Context… Stop
Stop Conditions
The user may check and use any of the listed stop conditions to terminate the analysis automatically. In
addition, more than one condition can be checked in which case the analysis will stop as soon as one of
the selected multiple conditions are met.
…when particle count reaches [user set value] particles: The analysis (sample run) ends when the
specified number of particles have been reached.
…when run time exceeds [user set value] minutes [user set value] seconds: The analysis ends when
the specified time limit is reached.
…after [user set value] collage files: The analysis ends when the specified of collage files have been
collected.
…after [user set value] camera images: The analysis ends when the specified number of camera
images have been collected.
…after [user set value] mL of fluid imaged: The analysis ends when the specified amount of sample
(in mL) has been imaged.
…only when sample volume is aspirated or operator terminates analysis: The experiment ends
when the sample volume entered on the Fluidics Tab is aspirated or the user selects the Stop button. If
this stop criterion is selected, the remaining stop criteria are grayed out and no longer available.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluid Imaging Technologies FlowCAM and is the answer not in the manual?

Fluid Imaging Technologies FlowCAM Specifications

General IconGeneral
BrandFluid Imaging Technologies
ModelFlowCAM
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals