4-22
Note
At voltage outputs of 100 V and above (nominal), you may notice a slight
high-pitched sound. This is normal.
5. Press a multiplier key, if necessary. For example, press c.
6. Output in volts. Press V.
Output in dBm. Press b V. Select an impedance for dBm from a list on the
Control Display using the rightmost softkey.
7. The Control Display now shows the amplitude of your entry. For example,
2.44949 V (below).
nn073f.eps
8. Press the numeric keys and decimal point key to enter the desired frequency output
(maximum five numeric keys). Press a multiplier key, if necessary. For example,
press the kilo multiplier key K. Then press the H key. For example, 1.1234
kHz (below).
nn074f.eps
9. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
nn075f.eps
10. Press O to activate the calibrator output.
Several softkey labels appear on the Control Display in the ac voltage function,
depending on which waveform is selected: DUTY, OFFSET and WAVE.