8-64. Leveled Sine Wave Verification: Harmonics
Table 8-23. Leveled Sine Wave Verification: Harmonics
Harmonic
Nominal
Value (V p-p) Frequency
Measured
Value (dB)
Deviation
(dB)
1-Year Spec.
(dB)
2nd harmonic 0.0399 50 kHz -33
3rd+ harmonic 0.0399 50 kHz -38
2nd harmonic 0.099 50 kHz -33
3rd+ harmonic 0.099 50 kHz -38
2nd harmonic 0.399 50 kHz -33
3rd+ harmonic 0.399 50 kHz -38
2nd harmonic 1.2 50 kHz -33
3rd+ harmonic 1.2 50 kHz -38
2nd harmonic 5.5 50 kHz -33
3rd+ harmonic 5.5 50 kHz -38
2nd harmonic 5.5 100 kHz -33
3rd+ harmonic 5.5 100 kHz -38
2nd harmonic 5.5 200 kHz -33
3rd+ harmonic 5.5 200 kHz -38
2nd harmonic 5.5 400 kHz -33
3rd+ harmonic 5.5 400 kHz -38
2nd harmonic 5.5 800 kHz -33
3rd+ harmonic 5.5 800 kHz -38
2nd harmonic 5.5 1 MHz -33
3rd+ harmonic 5.5 1 MHz -38
2nd harmonic 5.5 2 MHz -33
3rd+ harmonic 5.5 2 MHz -38
2nd harmonic 5.5 4 MHz -33
3rd+ harmonic 5.5 4 MHz -38
2nd harmonic 5.5 8 MHz -33
3rd+ harmonic 5.5 8 MHz -38
2nd harmonic 5.5 10 MHz -33
3rd+ harmonic 5.5 10 MHz -38
2nd harmonic 5.5 20 MHz -33
3rd+ harmonic 5.5 20 MHz -38
2nd harmonic 5.5 40 MHz -33
3rd+ harmonic 5.5 40 MHz -38
2nd harmonic 5.5 80 MHz -33
3rd+ harmonic 5.5 80 MHz -38
2nd harmonic 5.5 100 MHz -33
3rd+ harmonic 5.5 100 MHz -38
2nd harmonic 5.5 200 MHz -33
3rd+ harmonic 5.5 200 MHz -38
2nd harmonic 5.5 400 MHz -33
3rd+ harmonic 5.5 400 MHz -38
2nd harmonic 5.5 600 MHz -33
3rd+ harmonic 5.5 600 MHz -38