5520A
Operators Manual
8-86
8-111. Verification Tables
Before the SC300 Option leaves the Fluke factory, it is verified to meet its specifications
at the following test points. The verification test points are provided here as a guide
when re-verification is desired.
8-112. Voltage Function Verification: AC Voltage into a 1 M
Ω
Load
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year
Spec. (mV)
5.0 mV 10 Hz 0.11
5.0 mV 100 Hz 0.11
5.0 mV 1 kHz 0.11
5.0 mV 5 kHz 0.11
5.0 mV 10 kHz 0.11
10.0 mV 10 kHz 0.12
20.0 mV 100 Hz 0.15
20.0 mV 1 kHz 0.15
20.0 mV 10 kHz 0.15
50.0 mV 10 kHz 0.23
89.0 mV 10 Hz 0.32
89.0 mV 10 kHz 0.32
100.0 mV 10 kHz 0.35
200.0 mV 100 Hz 0.60
200.0 mV 1 kHz 0.60
200.0 mV 10 kHz 0.60
500.0 mV 10 kHz 1.35
890.0 mV 10 Hz 2.32
890.0 mV 10 kHz 2.32
1.0 V 100 Hz 2.60
1.0 V 1 kHz 2.60
1.0 V 10 kHz 2.60
2.0 V 10 kHz 5.10
5.0 V 10 Hz 12.60
5.0 V 10 kHz 12.60
10.0 V 10 kHz 25.10
20.0 V 10 kHz 50.10