5-55. Programming the ISR, ISCR, and ISCE
To read the contents of the ISR, send the remote command, ISR?. To read the contents
of the ISCR0 or 1, send the remote command, ISCR0?, or ISCR1?. To read the
contents of the ISCE0 or 1, send the remote command, ISCE0?, or ISCE1?. The
Calibrator responds by sending a decimal number that represents bits 0 through 15.
Every time you read the ISCR0 or 1, its contents are zeroed. The following sample
program reads all five registers:
10 ! THIS PROGRAM READS THE ISR, ISCR, AND ISCE REGISTERS
20 ! NOTE THAT THE ICSR? COMMANDS CLEAR THE ISCR CONTENTS
30 PRINT @6, ISR? ! ASK ISR CONTENTS
40 INPUT @6,A% ! RETRIEVE REGISTER CONTENTS FROM 5520A
50 PRINT @6, ISCR0? ! ASK FOR AND CLEAR ISCR0 CONTENTS
60 INPUT @6, B% ! RETRIEVE REGISTER CONTENTS FROM 5520A
70 PRINT @6, ISCE0? ! ASK FOR ISCE0 CONTENTS
80 INPUT @6, C% ! RETRIEVE REGISTER CONTENTS FROM 5520A
50 PRINT @6, ISCR1? ! ASK FOR AND CLEAR ISCR1 CONTENTS
60 INPUT @6, D% ! RETRIEVE REGISTER CONTENTS FROM 5520A
70 PRINT @6, ISCE1? ! ASK FOR ISCE1 CONTENTS
80 INPUT @6, E% ! RETRIEVE REGISTER CONTENTS FROM 5520A
90 PRINT ISR = ;A% ! DISPLAY ISR
100 PRINT ISCR0 = ;B% ! DISPLAY ISCR0
110 PRINT ISCE0 = ;C% ! DISPLAY ISCE0
100 PRINT ISCR1 = ;D% ! DISPLAY ISCR1
110 PRINT ISCE1 = ;E% ! DISPLAY ISCE1
120 END
Convert the returned variables into binary, and you can read the status of the instrument.
For example if a register contains 128, its binary equivalent is: 00000000 10000000.
Therefore, bit 7 (HIVOLT) is set (1) and the rest of the bits are reset (0).
By setting the bits in an ISCE register, you can mask (disable) the associated bits in the
ISCR. For example, to cause an SRQ interrupt when the output has settled, bit 12
(SETTLED) in the ISCE1 register must be 1. (The ISCB bit must also be enabled in the
SRE.) The following sample program loads a decimal 1024 into the ISCE, which sets bit
12 and resets the other bits:
10 ! THIS PROGRAM LOADS 00010000 00000000 BINARY INTO THE ISCE
20 PRINT @6, ISCE 4096 ! LOAD DECIMAL 4096 INTO ISCE
30 PRINT @6, ISCE? ! READ BACK ISCE VALUE
40 INPUT @6, A% !
50 PRINT ISCE = ;A% ! PRINT IT, IT SHOULD BE 4096
60 END
5-56. Output Queue
The output queue is loaded whenever a query is processed, and holds up to 800
characters. The controller reads it with a statement such as a BASIC INPUT statement,
removing what it reads form the queue. If the queue is empty, the Calibrator does not
respond to the INPUT statement from the controller. The Message Available (MAV) bit
in the Serial Poll Status Byte is 1 if there is something in the output queue and 0 if the
output queue is empty.