5520A-SC300 Option
Oscilloscope Calibration Option Specifications
8
8-61
8-79. Edge Function Specifications
Edge Characteristics into 50
1-Year Absolute
Uncertainty,
tcal ± 5°C
Amplitude
Range (p-p) 4.5 mV to 2.75 V ± (2% of output + 200 µV)
Resolution 4 digits
Adjustment Range ±10% around each sequence value
(indicated below)
Sequence 5 mV, 10 mV, 25 mV, 50 mV,
100 mV, 250 mV, 500 mV, 1 V, 2.5 V
Other Edge Characteristics
Frequency Range 1 kHz to 1 MHz ± (25 ppm of setting + 15
mHz)
Rise Time < 400 ps
Leading Edge
Aberrations
within 10 ns < (3% of output + 2 mV)
10 to 30 ns < (1% of output + 2 mV)
after 30 ns < (0.5% of output + 2 mV)
Typical Duty Cycle 45% to 55%