8-90
8-116. Edge Function Verification
Nominal Value (p-p) Frequency
Pulse Response Time
(ηs)
1-Year Spec.
(ps)
25.0 mV 1 MHz 400
250.0 mV 1 MHz 400
250.0 mV 10 kHz 400
250.0 mV 100 kHz 400
250.0 mV 1 MHz 400
2.5 V 1 MHz 400
8-117. Wave Generator Function Verification: 1 M
Load
Waveform
Nominal
Value (p-p) Frequency
Measured Value
(p-p)
Deviation
(mV)
1-Year
Spec. (mV)
Square 5.0 mV 10 kHz 0.25 mV
Square 20.0 mV 10 kHz 0.70 mV
Square 89.0 mV 10 kHz 2.77 mV
Square 219.0 mV 10 kHz 6.67 mV
Square 890.0 mV 10 kHz 26.80 mV
Square 6.5 V 10 kHz 195.10 mV
Square 55.0 V 10 kHz 1.65 V
Sine 5.0 mV 10 kHz 0.25 mV
Sine 20.0 mV 10 kHz 0.70 mV
Sine 89.0 mV 10 kHz 2.77 mV
Sine 219.0 mV 10 kHz 6.67 mV
Sine 890.0 mV 10 kHz 26.80 mV
Sine 6.5 V 10 kHz 195.10 mV
Sine 55.0 V 10 kHz 1.65 V
Triangle 5.0 mV 10 kHz 0.25 mV
Triangle 20.0 mV 10 kHz 0.70 mV
Triangle 89.0 mV 10 kHz 2.77 mV
Triangle 219.0 mV 10 kHz 6.67 mV
Triangle 890.0 mV 10 kHz 26.80 mV
Triangle 6.5 V 10 kHz 195.10 mV
Triangle 55.0 V 10 kHz 1.65 V