8-122. Marker Generator Function Verification
Nominal Interval Measured Interval Deviation 1-Year Spec.
5s 25.12 ms
2.00s 4.05 ms
1s 1.03 ms
500.00 ms 262.50 µs
200.00 ms 45.00 µs
100.00 ms 12.50 µs
50.00 ms 3.75 µs
20.00 ms 900.000 ns
10.00 ms 350.00 ns
5.00 ms 150.00 ns
2.00 ms 54.000 ns
1.00 ms 26.000 ns
500.00 µs 12.750 ns
200.00 µs 5.040 ns
100.00 µs 2.510 ns
50.00 µs 1.287 ns
20.00 µs 0.506 ns
10.00 µs 0.252 ns
5.00 µs 0.125 ns
2.00 µs 0.050 ns
1.00 µs 0.025 ns
500.000 ns 0.013 ns
200.000 ns 5.000 ps
100.000 ns 2.500 ps
50.000 ns 1.250 ps
20.000 ns 0.500 ps
10.000 ns 0.250 ps
5.000 ns 0.125 ps
2.000 ns 0.050 ps