5520A-SC600 Option
Verification Tables
8
8-45
8-62. Leveled Sine Wave Verification: Amplitude
Table 8-21. Leveled Sine Wave Verification: Amplitude
Nominal
Value (V p-p)
Frequency Measured
Value (V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
0.005 50 kHz 0.0004
0.0075 50 kHz 0.00045
0.0099 50 kHz 0.000498
0.01 50 kHz 0.0005
0.025 50 kHz 0.0008
0.039 50 kHz 0.00108
0.04 50 kHz 0.0011
0.07 50 kHz 0.0017
0.099 50 kHz 0.00228
0.1 50 kHz 0.0023
0.25 50 kHz 0.0053
0.399 50 kHz 0.00828
0.4 50 kHz 0.0083
0.8 50 kHz 0.0163
1.2 50 kHz 0.0243
1.3 50 kHz 0.0263
3.4 50 kHz 0.0683
5.5 50 kHz 0.1103
8-63. Leveled Sine Wave Verification: Frequency
Table 8-22. Leveled Sine Wave Verification: Frequency
Nominal
Value (V p-p) Frequency
Measured
Value (Hz)
Deviation
(Hz)
1-Year Spec.
(Hz)
5.5 50 kHz 0.125
5.5 500 kHz 1.25
5.5 5 MHz 12.5
5.5 50 MHz 125
5.5 500 MHz 1250