5520A-SC300 Option
Oscilloscope Calibration Option Specifications
8
8-63
8-81. Time Marker Function Specifications
Time Marker into 50 Ω 5 s to 100 µs 50 µs to 2 µs
1 µs to
20 ns
10 ns to
2 ns
1-Year Absolute
Uncertainty, tcal ± 5°C
± (25 + t *1000)
ppm [1]
± (25 + t *15,000)
ppm [1]
± 25 ppm ± 25 ppm
Wave Shape pulsed
sawtooth
pulsed sawtooth pulsed
sawtooth
sine
Typical Output Level > 1 V pk > 1 V pk > 1 V pk > 2 V p-p [2]
Sequence 5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms)
Adjustment Range At least ± 10% around each sequence value indicated above.
Resolution 4 digits
[1] t is the time in seconds.
[2] The 2 ns time marker is typically > 0.5 V p-p.
8-82. Wave Generator Specifications
Wave Generator Characteristics
Square Wave, Sine Wave, and Triangle Wave
into 50 Ω or 1 MΩ
Amplitude
Range into 1 MΩ: 1.8 mV to 55 V p-p
into 50 Ω: 1.8 mV to 2.2 V p-p
1-Year Absolute Uncertainty, tcal ±
5 °C,
10 Hz to 10 kHz
± (3% of p-p output + 100 µV)
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Typical dc Offset Range 0 to ± (≥40% of p-p amplitude) [1]
Frequency
Range 10 Hz to 100 kHz
Resolution 4 or 5 digits depending upon frequency
1-Year Absolute Uncertainty, tcal ± 5°C ± (25 ppm + 15 mHz)
[1]The DC offset plus the wave signal must not exceed 30 V rms.