1-37. AC Current, Square Wave Characteristics (typical)
Range LCOMP Risetime Settling Time Overshoot
I < 6 A @ 400 Hz off 25 µs 40 µs to 1% of final value < 10% for < 1 V Compliance
3 A & 20 A Ranges on 100 µs 200 µs to 1% of final value < 10% for < 1 V Compliance
1-38. AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz Aberrations
0.3% of p-p value, from 10% to 90% point < 1% of p-p value, with amplitude > 50% of range