The *OPC command is similar in operation to the *OPC? query, except that it sets bit 0
(OPC for “Operation Complete”) in the Event Status Register to 1 rather than sending a
1 to the output buffer. One simple use for *OPC is to include it in the program in order
for it to generate an SRQ (Service Request). Then an SRQ handler written into the
program can detect the operation complete condition and respond appropriately. You can
use *OPC similarly to *OPC?, except your program must read the ESR to detect the
completion of all operations. The following sample program shows how you can use
*OPC.
10 REMOTE
20 PRINT @4, OUT 100V,1KHZ;OPER;*OPC ! 5520A ADDRESS IS 4
30 PRINT @4, *ESR? ! PUT THE ESR BYTE IN BUFFER
40 INPUT @4, A% ! READ THE ESR BYTE
50 IF (A% AND 1%) = 0% GOTO 30 ! TRY AGAIN IF NO OPC
60 PRINT OUTPUT SETTLED
70 END
The *WAI command causes the Calibrator to wait until any prior commands have been
completed before continuing on to the next command, and takes no other action. Using
*WAI is a convenient way to halt operation until the command or commands preceding
it have completed. The following sample program shows how you can use *WAI.
10 REMOTE
20 PRINT @4, OUT 100V,1KHZ;OPER;*WAI ! 5520A ADDRESS IS 4
30 PRINT @4, OUT? ! READ THE OUTPUT VALUE
40 PRINT @4, A$,B$,C$ ! A$ CONTAINS THE OUTPUT VALUE
50 PRINT OUTPUT SETTLED
60 PRINT OUTPUT IS: ;A$;B$; at ;C$
70 END
5-64. Taking a Thermocouple Measurement
The following program takes one temperature measurement at a time.
10 REM Set Bus Timeout to 20 seconds, Init IEEE Bus
20 TIMEOUT 20 * 1000
30 INIT PORT 0
40 CLEAR @6
100 REM Reset 5520A, TC measurement mode
110 PRINT @6,*RST; TC_TYPE J; TC_MEAS FAR
200 PRINT Hit Carriage Return to take a Reading
210 INPUTLINE A$
220 REM Request the measurement value
230 PRINT @6, VAL?
240 REM Read measurement, unit
250 INPUT @6, M,U$
260 GOTO 200
5-65. Taking a Pressure Measurement
The following program takes one pressure measurement at a time.
10 REM Set Bus Timeout to 20 seconds, Init IEEE Bus
20 TIMEOUT 20 * 1000
30 INIT PORT 0
40 CLEAR @6
100 REM Reset 5520A, pressure measurement mode
110 PRINT @6,*RST; PRES_MEAS
200 PRINT Hit Carriage Return to take a Reading
210 INPUTLINE A$
220 REM Request the measurement value
230 PRINT @6, VAL?
240 REM Read measurement, unit
250 INPUT @6, M,U$
260 GOTO 200