5520A-SC600 Option
Verification Tables
8
8-41
8-56. Verification Tables
The verification test points areprovided here as a guide when verification to one-year
specifications is desired.
8-57. DC Voltage Verification
Table 8-16. DC Voltage Verification
(1 M output impedance unless noted)
Nominal
Value (V dc)
Measured Value
(V dc)
Deviation
(V dc) 1-Year Spec. (V dc)
0 0.00004
0.00125 0.000040625
-0.00125 0.000040625
0.00249 0.000041245
-0.00249 0.000041245
0.0025 0.00004125
-0.0025 0.00004125
0.00625 0.000043125
-0.00625 0.000043125
0.0099 0.00004495
-0.0099 0.00004495
0.01 0.000045
-0.01 0.000045
0.0175 0.00004875
-0.0175 0.00004875
0.0249 0.00005245
-0.0249 0.00005245
0.025 0.0000525
-0.025 0.0000525
0.0675 0.00007375
-0.0675 0.00007375
0.1099 0.00009495
-0.1099 0.00009495
0.11 0.000095
-0.11 0.000095
0.305 0.0001925
-0.305 0.0001925
0.499 0.0002895
-0.499 0.0002895
0.5 0.00029
-0.5 0.00029
1.35 0.000715
-1.35 0.000715
2.19 0.001135
-2.19 0.001135
2.2 0.00114
-2.2 0.00114
6.6 0.00334
-6.6 0.00334