Front Panel Operation
RMS Versus p-p Amplitude
4
4-17
HI
LO
TRIG
GUARD
TC
20A
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
CHART RECORDER INPUT
nn048f.eps
Figure 4-9. UUT Connection: Temperature (RTD)
HI
LO
TRIG
GUARD
TC
20A
NORMAL AUX
SCOPE
OUT
V, , ,
RTD
A, -SENSE,
AUX V
20V PK MAX
20V PK MAX
5520A
CALIBRATOR
!
60V
24V
MAX
OFFSET
ON/OFF
F/C
HOLD
K/J THERMOMETER
51
Connection wiring must match thermocouple type, e.g., K, J, etc.
nn049f.eps
Figure 4-10. UUT Connection: Temperature (Thermocouple)
4-19. RMS Versus p-p Amplitude
The 5520A Calibrator ranges for sinusoidal ac functions are specified in rms (root-mean-
square; the effective value of the wave form). For example, 1.0 to 32.999 mV, 33 to
329.999 mV, 0.33 to 3.29999 V and so forth. The sine wave outputs are in rms, while the
triangle wave, square wave, and truncated sine wave outputs are in p-p. The relationship
between p-p and rms for the non-sine wave types are as follows:
Square wave p-p x 0.5000000 = rms
Triangle wave p-p x 0.2886751 = rms
Truncated Sine wave p-p x 0.2165063 = rms