Front Panel Operation
Waveform Types
4
4-47
4-36. Sine wave
When the wave selection is sine, a sine wave current or voltage signal is present on the
calibrator outputs (Figure 4-11). The variables for the sine wave are amplitude,
frequency, and dc offset voltage.
Peak
Period
RMS (70% Peak)
nn026f.eps
Figure 4-11. Sine wave
4-37. Triangle wave
When the wave selection is tri, the triangle wave is present on the calibrator outputs
(Figure 4-12). The variables for the triangle wave are amplitude, frequency, and dc offset
voltage. Whenever a triangle wave is selected, the Output Display indicates amplitudes
in p-p units.
Peak to Peak
nn027f.eps
Figure 4-12. Triangle Wave
4-38. Square Wave
When the wave selection is square, a square wave current or voltage signal is present on
the calibrator outputs (Figure 4-13). The variables for the square wave are duty cycle,
amplitude, frequency, and dc offset voltage. Whenever a square wave is selected, the
Output Display indicates amplitude in p-p units. If the calibrator is set for a single
voltage or current output, the duty cycle of the signal can be set through the keypad. To
enter a new duty cycle, press the DUTY CYCLE softkey and up to five numeric keys
followed by E. The negative-going edge of the square wave will move based on
the duty cycle setting.