Operators Manual
5-8
5-3. IEEE-488 Port Setup Procedure
Complete the following procedure to set up the Calibrator for remote operations using
the IEEE-488 remote control port. The purpose is to select GPIB as the interface and to
select the GPIB address for the interface.
1. Turn the Calibrator power on. You may operate the Calibrator during warmup, but
specifications are not guaranteed until warmup is complete.
2. Press S on the Calibrator front panel.
3. Negotiate the softkey selections shown below. Verify the HOST port selection is
gpib. Select the desired GPIB port address (0 to 30) using the UP/DOWN softkeys.
The factory default is 4.
Select
nn120f.eps
4. Press P (not E ) several times until the message STORE
CHANGES/DISCARD CHANGES appears or, if there were no changes, the reset
display. If you select STORE CHANGES, the gpib and host port setting are saved in
the instrument non-volatile memory.