Introduction and Specifications
Specifications
1
1-21
1-21. Temperature Calibration (RTD) Specifications
RTD
Range
°C
Absolute
Uncertainty
tcal ±5 °C
± °C [2] RTD
Range
°C
Absolute
Uncertainty
tcal ±5 °C
± °C [2]
Type [1]
90 days 1 year
Type [1]
90 days 1 year
-200 to -80 0.04 0.05 -200 to -80 0.03 0.04
-80 to 0 0.05 0.05
-80 to 0 0.04 0.05
0 to 100 0.07 0.07
0 to 100 0.05 0.05
100 to 300 0.08 0.09
100 to 260 0.06 0.06
300 to 400 0.09 0.10
260 to 300 0.07 0.08
400 to 630 0.10 0.12
300 to 400 0.07 0.08
Pt 395,
100
630 to 800 0.21 0.23
400 to 600 0.08 0.09
-200 to -80 0.04 0.05
Pt 385,
500
600 to 630 0.09 0.11
-80 to 0 0.05 0.05
-200 to -80 0.03 0.03
0 to 100 0.07 0.07
-80 to 0 0.03 0.03
100 to 300 0.08 0.09
0 to 100 0.03 0.04
300 to 400 0.09 0.10
100 to 260 0.04 0.05
Pt 3926,
100
400 to 630 0.10 0.12
260 to 300 0.05 0.06
-200 to -190 0.25 0.25 300 to 400 0.05 0.07
-190 to -80 0.04 0.04 400 to 600 0.06 0.07
-80 to 0 0.05 0.05
Pt 385,
1000
600 to 630 0.22 0.23
0 to 100 0.06 0.06 PtNi 385, -80 to 0 0.06 0.08
100 to 260 0.06 0.07 120
0 to 100 0.07 0.08
260 to 300 0.07 0.08 (Ni120)
100 to 260 0.13 0.14
300 to 400 0.08 0.09 Cu 427,
-100 to 260 0.3 0.3
400 to 600 0.08 0.10 10 [3]
Pt 3916,
100
600 to 630 0.21 0.23
-200 to -80 0.03 0.04
-80 to 0 0.03 0.04
0 to 100 0.04 0.04
100 to 260 0.04 0.05
260 to 300 0.11 0.12
300 to 400 0.12 0.13
400 to 600 0.12 0.14
Pt 385,
200
600 to 630 0.14 0.16
[1] Resolution is 0.003 °C
[2] Applies for COMP OFF (to the 5520A Calibrator front panel NORMAL terminals) and 2- wire and
4-wire compensation.
[3] Based on MINCO Application Aid No. 18