Remote Commands
Command Summary by Function
6
6-9
6-12. Status Commands
ERR? Returns the first error code with an explanation contained in the Calibrator error
queue, then removes that error code from the queue.
EXPLAIN? Explains an error code. This command returns a string that explains the error code
furnished as the parameter.
FAULT? Returns the first error code contained in the Calibrator error queue, then removes
that error from the queue.
FUNC? Returns the present output, measurement, or calibration function.
ISCE Loads two bytes into both the Instrument Status 1 to 0 Change Enable register and
the Instrument Status 0 to 1 Change Enable register.
ISCE? Returns the OR of the contents of the Instrument Status 1 to 0 Change Enable
register and the Instrument Status 0 to 1 Change Enable register.
ISCE0 Loads two bytes into the Instrument Status 1 to 0 Change Enable register.
ISCE0? Returns the contents of the Instrument Status 1 to 0 Change Enable register.
ISCE1 Loads two bytes into the Instrument Status 0 to 1 Change Enable register.
ISCE1? Returns the contents of the Instrument Status 0 to 1 Change Enable register.
ISCR? Returns the OR of the contents of the Instrument Status 1 to 0 Change Register
and the Instrument Status 0 to 1 Change Register and clears both registers.
ISCR0? Returns and clears the contents of the Instrument Status 1 to 1 Change Register.
ISCR1? Returns and clears the contents of the Instrument Status 0 to 0 Change Register.
ISR? Returns the contents of the Instrument Status Register.
ONTIME? Returns the time since the Calibrator was powered up last.
6-13. Thermocouple (TC) Measurement Commands
TC_MEAS Changes the operating mode to thermocouple measurement.
TC_OFFSET Sets a temperature offset for the thermocouple measurement mode.
TC_OFFSET? Returns the temperature offset when in the thermocouple measurement mode.
TC_OTCD? Returns whether or not the open thermocouple detection circuit is set.
TC_OTCD Activates or deactivates the open thermocouple detection circuit in thermocouple
measurement mode.
VAL? Returns the last thermocouple, pressure, or, for the -SC600, impedance
measurement value.
VVAL? Returns the last value of the thermocouple measurement in volts.