Fluke 5520A - Voltage Function Verification: DC Voltage into a 50 Ω Load

Fluke 5520A
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A
Operators Manual
8-88
Voltage Function Verification: AC Voltage into a 50
Load (cont.)
Nominal Value (p-p) Frequency Measured Value (p-p)
Deviation
(mV)
1-Year Spec.
(mV)
2.0 V 100 Hz 5.10
2.0 V 1 kHz 5.10
2.0 V 5 kHz 5.10
2.0 V 10 kHz 5.10
8-114. Voltage Function Verification: DC Voltage into a 50
Load
Nominal Value
(dc) Measured Value (dc) Deviation (mV)
1-Year Spec.
(mV)
0.0 mV 0.10
5.0 mV 0.11
-5.0 mV 0.11
10.0 mV 0.12
-10.0 mV 0.12
22.0 mV 0.15
-22.0 mV 0.15
25.0 mV 0.16
-25.0 mV 0.16
55.0 mV 0.24
-55.0 mV 0.24
100.0 mV 0.35
-100.0 mV 0.35
220.0 mV 0.65
-220.0 mV 0.65
250.0 mV 0.72
-250.0 mV 0.72
550.0 mV 1.47
-550.0 mV 1.47
700.0 mV 1.85
-700.0 mV 1.85
2.2 V 5.60
-2.2 V 5.60